Testset
Service Manual M695
09/93 3077-HWE Doc code 636 718
Functions of test box
All control voltages . . .
All supply- and control voltages between the control unit and the microscope
with XY-gear can be measured.
. . . to the microscope . . .
Control voltages to the microscope are measured relative to XY_GND, such
to the XY-gear relative to GND_XY.
. . . measure at low light
level
Unless otherwise indicated, set the lamp to minimum brightness. The voltages
indicated in Section 4 are only valid for this condition.
Voltage levels are falsfied by the voltage drop between the control unit and
the microscope. Because of the current taken by the lamp, this voltage drop
to PGND (measuring ground) can be as much as 0.9V. The control voltages
of focus and zoom are measured against PGND and are therefore influenced.