Important: With the illuminator on, insert the filterhol-
der with diaphragm into slide-in unit 3 to avoid being
suddenly exposed to excessively bright light (page 22).
Remove the binocular tube.
Loosen the 3 socket-head
screws.
Remove the cover.
Loosen the fastening
screw and take out the
filter carrier.
Insert the desired filter
carrier. Fit both of the
two guide pins in the
fluorescence module into
the holes on the filter
carrier.
Tighten the fastening
screw securely.
Replace the cover and
tighten the 3 socket-head
screws securely.
Replace the binocular tube.
Insert a suitable exciter filter (page 22).
Take out the filterholder with diaphragm.
Re-adjust the light source (page 15).
Stereo-fluorescence module
Focusing drive base of stand
Connect manually-operated or motor-driven focusing drive,
complete with column, to base, as described in user manual
M2-105-0EN.
Before using the motor focus, you must read
the safety directions in the accompanying user
manual M2-267-104.
For the Leica stereomicroscope models MS5, MZ6, MZ8, MZ12,
MZAPO with manual focusing drive, fit the micro scope in
the upper position on the focusing drive (refer to user manual for
stereomicroscope).
Fluorescence module stereomicroscope
• Assembleinthefollowingorderwithoutdouble-irisdia-
phragm: Optics carrier, fluorescence module, bin ocular tube
or phototube
• Assembleinthefollowingorderwithdouble-irisdiaphragm:
Optics carrier, double-iris diaphragm, fluorescence module,
bin ocu lar tube or phototube
Align the fluorescence module on the optics carrier so that
the adapter for the lamp housing points backwards at
about 45°
Continue assembling in accordance with the user manual for
the stereomicroscope (see section “Fitting accessory tubes”)
Video-/phototubes
The fluorescence illumina tion is directed along the left-hand
beam path and could cause hot-spots in the image. Therefore:
PositiontheHUphototubeovertheright-handbeampath.
Changing the filter set
Removable filter sets that can hold various filters are available for
the fluorescence module. A filter set consists of a filter carrier
with barrier filter and of a filterholder with excitation filter.
If you wish to change the filter set while you are in the middle
of a task, leave the illuminator switched on so that you will
not have to wait for the lamp to cool (page 12) before continuing.