Leica M Stereomicroscopes – Assembly
Stand
Incident-light stand
Remove the stage plate.
Insert 3 Allen screws into the underside
of the baseplate and secure them to the
bottom end of the side-faced column.
Insert the stage plate.
To fit the microscope carrier and the stereomicroscope,
see page 34
Transmitted-light stand
Pull the power cable out of the socket in the stand.
Remove the glass stage plate.
Carefully invert the transmitted-
light base and remove 7 Allen
screws from the base.
Lift off the lower part of the base.
Reverse the upper part of the base.
Insert 3 Allen screws into the cover
plate from beneath and secure
them to the bottom end of the side-
faced column.
Replace and secure the base.
Insert the glass stage plate.
Fit the microscope carrier and the
stereomicroscope as described
on page 34.
Technical data see page 74