Operating Manual V 1.1 – 05/2007
Fig. 56
Bacterial filter
Bactericidal filter 350/5865, pack of 1 unit. Rec-
ommendation: replace bactericidal filter every
three months.
(note date on filter when installing it)
Anti-roll guide system
Anti-roll plate (with glass stage plate)
Available with various spacers:
70 mm - 50 μm, for section thickness: < 20 μm
70 mm -100 μm, for section thickness: 20 μm -
100 μm
70 mm -150 μm, for section thickness: > 100 μm
The 50 μm and 100 μm anti-roll plates
are included in the standard scope of
delivery of the blade holder CE.
70mm
11. Optional accessories
Fig. 55
Fig. 54
Straightedge (blade rest)
Insert for low-profile blades for blade holder CE
(0477 43005) - replacement
The blade rest is also included in the
standard scope of delivery of the
blade holder CE.
Filter bags and bacteria filter must be
disposed of according to applicable
laboratory regulations for infectious
material. Filters must be replaced, not
cleaned.
When using low-profile blades, the blade rest (7)
must be inserted into the blade holder first, then
the low-profile blade (see page 50, Fig. 40).