Leica DM750 P Manual 28
Setting up Polarized Light (cont.)
The Analyzer is located in the module
between the viewing tube and the stand
either in a slider module or in an Analyzer
Bertrand Lens module. If you have a slider, slide
the Analyzer into position.
3. If you have an Analyzer/Bertrand Lens
module, swing in the Analyzer by flipping
the lower rocker of the module to the
right (the "A" position/Analyzer in), while
ensuring the upper rocker of the Analyzer/
Bertrand Lens Module is flipped to the left
(the "O" position/Bertrand Lens out).
4. Increase the illumination intensity slightly.
You will notice while looking into the
microscope with the Polarizer and Analyzer
in place and no specimen, that the field will
look dark. If not, rotate the polarizer slightly
until you have the darkest field of view.