Page view 24
Leica M series Manual 25
Preparations
Hold the objectives firmly during
assembly and disassembly so that they
do not fall onto the stage plate.
Move the drive housing all the way upwards
and remove the optics carrier, if the carrier
has already been installed.
Assembly
1. Remove the transport anchor from the
objective nosepiece.
2. Rotate the moving part by 90° and attach
the objective nosepiece to the drive hous-
ing from the front (!). Screw the objective
nosepiece firmly into place.
3. Unscrew the three Phillips screws on the
objective mount of the optics carrier and
remove the intermediate ring.
4. Screw the optics carrier onto the objective
nosepiece.
5. Screw both objectives onto the objective
nosepiece. It makes no difference which
position an objective occupies.
6. Unscrew the locking screws on both sides
of the objective nosepiece.
You can now adjust the parfocality (see instruc-
tions on the next page).